Standard Incubator vs. Shaking Incubator — Differences and Application Guide

November 02, 2026

1. Two Branches of Incubation: Static Culture and Shaking Culture


A standard incubator holds the internal temperature (and, depending on the model, the CO₂ concentration) at a set value while samples stay motionless. This is static culture. Liquid media cultured this way get no forced mixing, so oxygen transfer and mixing are limited. For fast growth of oxygen-demanding microorganisms in liquid media, a shaking incubator is the better choice.

 

A shaking incubator adds an orbital shaking function to the temperature control of a standard incubator. A motor with an eccentric drive moves the internal platform in a circular or elliptical orbit, so the liquid medium in flasks and other culture vessels is mixed continuously. 

 

Shaking keeps the culture fluid moving, which promotes mass transfer at the gas–liquid interface and raises the rate at which oxygen enters the medium. It also reduces concentration gradients of nutrients and metabolites and helps keep cells and microorganisms from settling. Holding shaking speed, orbital diameter and medium volume constant also standardizes operating conditions across repeated experiments.

2. Key Comparison


The core difference is motion: a standard incubator only holds conditions steady, while a shaking incubator also mixes the culture.

 

Item 

 Standard Incubator

Shaking Incubator 

Core Function

Maintains a constant temperature 

Temperature control + orbital shaking

Sample State

Static 

Continuous orbital shaking 

Liquid Culture

Possible 

Advantageous when mixing and oxygen transfer are needed 

Typical Uses

Microbial culture, culture testing, constant temperature tests

Liquid microbial culture, suspension cell culture

Typical Vessels

Petri dishes, plates, flasks

Erlenmeyer flasks, shaking speed (rpm), orbital diameter 

Key Settings 

Temperature 

Temperature, shaking speed (rpm), orbital diameter 

CO₂ / Humidity 

Configured separately depending on instrument type

Depends on model (e.g., C5, C5H)

Selection Focus

Stable constant temperature environment

Mixing, oxygen transfer, keeping cells in suspension



3. Typical Applications of Shaking Incubators


Shaking incubators are used wherever a liquid culture needs continuous mixing and oxygen supply.

 
Broth culture of microorganisms. Aerobic bacteria and yeast are grown in liquid media to reach high biomass quickly. Shaking is widely used here for oxygen transfer and mixing.


Suspension animal cell culture. Research on CHO cells and hybridoma for antibody and recombinant protein production, and screening of culture conditions, are typical examples.


CO₂ and humidity control. Bicarbonate-buffered media may require CO₂ control. When evaporation and osmolality changes during long runs matter, a humidity-controlled (H) model can also be considered.


Production of secreted compounds. Experiments in which microorganisms secrete enzymes, antibiotics or organic acids during culture.


Optimization and scale-up. Optimizing culture conditions and running the stage before bioreactor scale-up.


Environmental and food microbial QC. Growing microorganisms from a sample in liquid medium for quantitative and qualitative analysis.


For adherent cells that must be expanded in large quantity over a wide culture surface, a roller culture device is another option. The configuration of the SH roller culture system is covered in a separate guide.

 

 

4. SH Scientific Shaking Incubators (IS Series)


The IS series covers benchtop, high-capacity, CO₂/humidity-controlled, and vertical or top-loading models, so you can match the instrument to experiment scale and purpose.

 

4.1 Benchtop (small laboratories)


The IS-RDD3 has a temperature range of ambient −15 °C to 60 °C (minimum 4 °C), accuracy of 0.1 °C and uniformity of ±0.5 °C (at 37 °C). Shaking speed is 30–400 rpm (±1 rpm) with a 26 mm orbital diameter. The single-tier platform measures 450 × 400 mm and holds up to 20 flasks of 250 mL.

 

A curved glass door lets you observe samples without opening the chamber, and an automatic load-balancing system is supported. For tighter spaces, the ultra-compact IS-RSDA, which holds two 1 L flasks, is also worth considering.

 

4.2 High-capacity (mid-size and large laboratories, industry)

 

The IS-RDS4 has a temperature range of ambient −18 °C to 60 °C (minimum 4 °C), accuracy of ±0.1 °C and uniformity of ±0.3 °C (at 37 °C). Shaking speed is 0 and 30–300 rpm (±1 rpm) with a 26 mm orbital diameter. Internal volume is 250 L with up to three tiers, a 770 × 470 mm platform and a maximum load of 20 kg. It has a 10-inch color touchscreen, 12 storable programs, a USB data logger, and humidity display and alarm.

 

The IS-RDS6 series (401 L class) suits mid-size laboratories and pilot-scale work that need many flasks at once. Its temperature uniformity is ±0.5 °C (at 37 °C), somewhat wider than the IS-RDS4. Its temperature range, shaking speed, orbital diameter and touchscreen follow the same pattern as the IS-RDS4.

 

Both models use a Triple Eccentric Drive. The platform is supported at three points by eccentric shafts (cams) at the same time, which keeps the orbital motion stable on large platforms under heavy loads. Noise is 55 dB or less for the IS-RDS4 and 53 dB or less for the IS-RDS6, both measured in refrigerated operation.

 

Model Family 

CO​2 ​Control 

Humidity Control

Orbital Diameter 

Internal Volume 

Flasks (1,000 mL / 250 mL) 

IS-AX

No

No

26 mm 

190 L

12 / 30 

IS-RDS4 / RDS4T

No

No

26 mm 

250 L

15 / 40 

IS-RDS4C5 / RDS4C5T

Yes

No

50 mm 

250 L

15 / 40 

IS-RDS4C5H

Yes

Yes

50 mm 

250 L 

15 / 40 

IS-RDS6 / RDS6T

No

No

26 mm 

401 L

28 / 77 

IS-RDS6C5 / RDS6C5T

Yes

No

50 mm 

401 L

28 / 77

IS-RDS6C5H

Yes

Yes

50 mm 

401 L

28 / 77 

 

C5 models add CO₂ control, and C5H models add both CO₂ and humidity control. C5 and C5H models share the internal volume and flask capacity of the base model with the same number (RDS4 or RDS6) and use a 50 mm orbital diameter. Consider CO₂ control for suspension animal cells in bicarbonate-buffered media, and humidity control (H) when evaporation and osmolality changes during long runs matter.

 

4.3 Do not compare shaking incubators by rpm alone


The movement of the medium is not set by rpm alone. At the same 150 rpm, mixing and oxygen transfer change with the orbital diameter (26 mm or 50 mm), the flask size (250 mL or 2 L) and the fill volume. When choosing an instrument, check shaking speed, orbital diameter, flask size and medium volume together. This is why the SH lineup includes both 26 mm and 50 mm orbital diameter models.

 

4.4 Vertical and top-loading models (space efficiency, large flasks)


IS-RDV1: two-tier vertical structure with UV sterilization.


IS-RDV3 and IS-RDV35: dual-door, large-capacity models for wide-orbit shaking.


IS-RDH1 and IS-RDH2: top-opening, top-loading models that hold up to 14 flasks of 2 L (IS-RDH1) or 6 flasks of 5 L (IS-RDH2). They suit laboratories that need large-volume culture in little space, or that run many large flasks.



5. Selection Criteria 


1 - Does the sample need to move during culture? If the goal is to keep samples static at a constant temperature, consider a standard incubator. If a liquid medium must be mixed continuously, consider a shaking incubator. Animal cell culture also requires checking CO₂, humidity and other environmental needs.


2 - What shaking speed (rpm) and orbital diameter do you need? Bacteria and yeast need different shaking conditions from animal cells, and rpm alone is not enough to decide. Consider shaking speed, orbital diameter, flask shape and medium volume together. Models for suspension animal cells may use a larger orbital diameter to get enough mixing at low shear.


3 - Is this cell culture that needs CO₂ or humidity control? Bicarbonate-buffered media require CO₂ control. In that case choose a C5 or C5H model (CO₂ only, or CO₂ and humidity).


4 - How many flasks, and of what size, will you run at once? Platform size and number of tiers decide the fit, from benchtop (IS-RDD3, IS-RSDA) to high-capacity (IS-RDS4, IS-RDS6).


5 - What are your space and power conditions? For limited space, also consider vertical and top-loading models (IS-RDV1, IS-RDH1, IS-RDH2).


6 - Do you need data logging or programmed operation? If you run many conditions repeatedly, check the number of storable programs and USB data logger support.


In short: confirm whether the sample must move, then choose among benchtop to high-capacity models by shaking speed and orbital diameter, CO₂ and humidity control, and the number of flasks you run at once. 

 

 

FAQ


Q: Can one shaking incubator also culture adherent cells in large quantity?

: Shaking incubators are optimized for liquid (suspension) culture. To produce vaccines or antibodies from adherent (monolayer) cells grown over a wide culture surface, consider a roller culture device or a roll-in incubator combination as well.

 

 

Glossary 


Static culture: culturing with the sample kept still, without physical movement.


Orbital shaking: shaking in which the platform does not rotate on its own. Every point on it moves in a circle of the same orbital diameter, in parallel.


Orbital diameter: the diameter of the circle the platform traces. At the same rpm, a different orbital diameter changes how the culture moves and how oxygen is transferred.


Triple Eccentric Drive: a drive that supports and rotates the platform in sync from eccentric shafts at three points, so it does not tilt under heavy loads or high speeds.


Suspension cells: cells that grow floating in the medium instead of attaching to the vessel surface. They are the main subject of shaking incubators.


Dissolved oxygen (DO): the amount of oxygen dissolved in the medium. Microorganisms and cells need it for metabolism, and shaking promotes its supply.

 

 

Explore Our Incubator Models

Browse Available models and detailed product specifications

Refrigerared Incubators - all models 

Shaking Incubators - all models 

SH Scientific Youtube Video / Shaking Incubator - IS-RDS6

 

Looking for the Right Incubator? 

Share your temperature requirements, sample type, and whether shaking is needed, and we will help you select the most suitable Refrigerated Incubator or Shaking Incubator for your application. 

Contact: martin@samheung21.com I jay@samheung21.com 