메뉴 바로가기 본문 바로가기

Articles

CVD Furnace vs. UHV Furnace - Differences in Vacuum Level, Pumping and Sealing

November 02, 2026

CVD furnaces and UHV (ultra-high vacuum) furnaces both carry the word "vacuum", so they are easy to mistake for equipment of a similar class. In fact they are designed for different purposes, with different sealing methods and exhaust system configurations. This guide compares the standard SH-CVD / SH-CVD-PK configuration with a UHV made-to-order case.

 

A CVD furnace (SH-CVD, SH-CVD-PK) is a catalog line that uses a process tube and gas sealing masks for gas-atmosphere heat treatment and basic CVD processes. A UHV furnace is a separate family (a made-to-order case) that combines a stainless steel tube, vacuum flanges and a cryopump to target a much lower base pressure.

 

 

 

1. Structural Comparison: CVD Furnace vs. UHV Furnace

 

The biggest differences are the sealing method and the exhaust system: a CVD furnace is built for gas atmospheres, while the UHV example adds the hardware needed to reach a far lower base pressure.

 

 Parameter

CVD Furnace 

(SH-CVD / SH-CVD-PK)

UHV Furnace 

(Made-to-order case)

Main Purpose

Gas-atmosphere heat treatment, basic CVD processes

Clean ultra-high vacuum heat treatment, surface research

Supply Form

Catalog line 

Many models by temperature and diameter

Made to order (custom build)

No standard catalog model 

Process Tube

1200 °C: Quartz tube 

1500–1800 °C: Alumina tube (by default; varies by model)

Stainless steel 316 (SS316)

Sealing Method

Gas sealing masks at both tube ends

Vacuum flanges (ISO200, KF16)

Viton vacuum gaskets, integrated cooling water ports ​

Roughing (Primary Pumping)

Oil rotary vacuum pump 

SH-V10, included from the CVD-PK stage

Oil Rotary vacuum pump 

960 L / min

High-Vacuum Pumping

None 

In the standard configuration 

8-inch cryopump 

Base Pressure

Configuration-dependent

Depends on sealing, piping, outgassing, etc.; must be confirmed separately  

5x10​-9​ Torr 

Value listed in the case's specification table

Heating Zones

Single zone or 3-zone split 

e.g., 200 mm x 3, diameter  Ø30 to Ø250

3-zone split 

300 mm x 3 = 900 mm; diameter Ø320

Maximum Operating Temperature

1200-1800°C 

Depending on model (furnace rated temperature); continuous operating temperature depends on tube material and must be checked separately

Maximum 800°C 

600°C in actual operation

 

The vacuum chamber of this UHV case combines an SS316 tube with vacuum flanges and Viton gaskets, and pairs an oil rotary pump with a cryopump. The system is designed for a base pressure of 5×10⁻⁹ Torr. That figure is a specification value from the case's vacuum system table. The pressure actually reached depends on the type of vacuum gauge, the measurement conditions and the condition of the chamber.

 

 

2. How Structure Shapes the Process


2.1 Tube material and sealing method


CVD and gas-flow furnaces are configured with a process tube, such as quartz or alumina, that suits the model and operating temperature. For example, the SH-CVD-PK 1200 °C class comes with a quartz tube as standard (alumina is optional), and the 1500–1800 °C class comes with an alumina tube as standard.

Quartz is widely used for its light transmission and thermal shock behavior, but the temperature at which it can be used continuously for long periods is limited. The rated maximum temperature of the furnace and the allowable temperature of the process tube you actually use must therefore be checked separately.

The sealing method also differs. A CVD furnace uses gas sealing masks that close both ends of the tube. The UHV case uses stainless steel vacuum flanges with cooling water ports, combined with Viton gaskets.

That configuration alone does not create ultra-high vacuum, however. A typical UHV system uses CF (ConFlat) flanges with metal gaskets, plus high-temperature bakeout to remove water and gas adsorbed on the chamber walls. Rubber gaskets such as Viton limit the bakeout temperature, which can affect the base pressure that can be reached.

The UHV case here still quotes an ultra-high vacuum specification by pairing Viton gaskets with an exhaust system that includes a cryopump. It is therefore not accurate to simplify this into "using a particular gasket material automatically gives ultra-high vacuum."

 

2.2 Exhaust system configuration


The standard vacuum configuration of the SH-CVD-PK compared in this guide pumps with an oil rotary pump only, and is built around gas atmosphere control and low-vacuum operation. The UHV case uses a two-stage setup: an oil rotary pump for primary pumping (roughing), followed by a cryopump for additional pumping.

A cryopump is a capture-type pump that condenses and adsorbs gas on a cryogenic surface to remove it. The pump itself uses no oil, and it is used to reach the high-vacuum to ultra-high-vacuum range.

 

2.3 Why base pressure differs: the role of residual gas


UHV is not needed simply because the number is lower. The lower the pressure, the less residual gas such as oxygen, water and hydrocarbons remains in the chamber, so the sample surface is less likely to react with that gas or become contaminated.

For this reason, research where surface and interface properties matter, and processes that are sensitive to residual gas, are often judged by residual gas conditions rather than by the pressure value itself.

UHV systems commonly use bakeout (heated degassing) to reduce the water and gas adsorbed on the chamber walls. The bakeout temperature that can actually be applied depends on the allowable temperature of components such as gaskets, gauges and valves, so it must be checked for each system.

The UHV case has cooling water ports on its flanges to separate the process temperature in the sample's hot zone from the temperature the sealing parts (flanges and gaskets) actually experience.

In a vacuum system, you need to manage not only the sample process temperature but also the temperature that vacuum components such as flanges, gaskets, gauges and pumps actually see. The case's conditions of 800 °C maximum and 600 °C in operation are set on top of this temperature management structure.

 

3. Applications and Use Cases


CVD furnaces (gas-flow packages and CVD turnkey packages) serve as base equipment for pyrolysis, annealing, oxygen-free atmosphere heat treatment, and CVD research that adds gas and vacuum configurations suited to the process conditions.

 

A UHV furnace can be considered for heat treatment, surface and interface research, and clean vacuum annealing where the influence of residual gases such as oxygen and water must be kept extremely low. This case has been applied to electrode material annealing and to vacuum brazing in aerospace and defense.

 

Whether UHV is actually needed is not decided by the type of sample alone. It should be judged from the base pressure and residual gas conditions the process requires.

 

Key takeaway: a CVD furnace is a catalog tool for gas-atmosphere heat treatment, made up of a process tube, gas sealing masks and a rotary pump. The UHV case is a made-to-order system designed for a far lower base pressure, using vacuum flanges and a cryopump.

 

Both carry the word "vacuum", but the orders of magnitude of the target base pressure are different. Confirm the pressure and residual gas conditions your process requires first, then decide which one you need.

 

 

4. Selection Criteria


Start from the pressure your process requires, then check the tube material, sealing method and pumping system of the matching line.

 

1 - Do you need gas-atmosphere heat treatment or basic CVD? Consider the gas-flow package (SH-CVD-*). It comes with a process tube, gas sealing masks and flow meters as standard

2 - Do you need an integrated CVD system that includes a vacuum pump and chiller? Consider the CVD turnkey package (SH-CVD-PK-*). Note that its vacuum pump is for gas purging and low-vacuum assist, and is not specified to guarantee ultra-high vacuum.


3 - Does your process actually require a base pressure in the 10⁻⁹ Torr range? For clean annealing and surface or interface research that must suppress residual gas effects to an extreme degree, consider a UHV furnace. What matters is the base pressure and residual gas conditions the process truly requires, not a particular material or process name, so define those first.


4 - Is the required vacuum level unclear? Define the target base pressure, gas type and tube diameter first. Then we recommend a pre-consultation to confirm whether the CVD line or the UHV line is the better fit.

 

In short: decide the pressure condition your process requires (a gas-atmosphere level, or an ultra-high vacuum with a base pressure in the 10⁻⁹ Torr range), then confirm the line whose tube material, sealing method and pumping system match it.

 

 

5. Photos of the made-to-order UHV tube furnace described in this guide

 


1. Overall view of the UHV tube furnace with the hinged heating chamber opened, showing the stainless steel process tube and the three-zone temperature controllers.

 


2. Close-up of the vacuum flange at the tube end, opened to show the tube opening and the water-cooling lines attached to the flange.

 


3. Cryopump mounted on the vacuum flange, with isolation valves and pneumatic lines for the pumping system.


Glossary


UHV (Ultra-High Vacuum): generally a very low pressure range of about 10⁻⁹ Torr or below. The exact boundary is defined differently by field and by reference.


Base pressure: the lowest pressure the system itself can reach, before any sample processing or gas introduction.


Process pressure: the pressure held during actual heat treatment or deposition, when gas is introduced or a reaction is run. It differs from base pressure. A low base pressure does not mean every process runs at that pressure.


Cryopump: a capture-type vacuum pump that condenses and adsorbs gas molecules on a cryogenically cooled surface. It is combined with pumps such as rotary pumps to reach high to ultra-high vacuum.


Gas sealing mask: a part that closes both ends of the process tube to seal it, and provides gas in/out ports for supplying and exhausting atmosphere gas.


Explore Our Furnace Models 

Browse available models and detailed product specifications

Gas-Flow Packages (SH-CVD) - All Models

CVD Turnkey Packages (SH-CVD-PK) - All Models


Looking for the Right Furnace? 

Share your process requirements, target temperature, and vacuum conditions, and we will help you find the most suitable CVD or UHV furnace solution.

Contact: martin@samheung21.com I jay@samheung21.com 

______________________________________________________________________________________________________________________________________

 

Written by Jay Bang, Bioengineer, SH Scientific